Tempus IC2
Integrated monitor and telemedicine solution
  • Product in Use Photograph
  • Standard Product Photograph

Tempus IC2

Integrated monitor and telemedicine solution
An advanced portable multi-parameter vital signs monitor with integrated telemedicine. Tempus IC2 allows trained non-medical users to manage medical emergencies in the air, at sea, and in remote locations. Along with built-in ease of use, the Tempus IC2 features clear visual guidance and automated data collection, helping users to feel confident in managing a medical emergency.

Features
Discreet access to medical infrastructure
No one can predict when or where they may be injured or taken ill, or how serious the situation may be. Tempus IC2 provides assistance to medical professionals through a reliable, secure communication link with voice and data automatically connecting via existing satellite or terrestrial communication systems, with no need for proprietary software or standalone equipment.
Secure connections for medical professionals
Connected medical support
Imagine being in a remote and challenging location and hours away from quality medical help. With Tempus IC2, users can securely connect 24/7 to medical professionals who can fully control Tempus IC2 remotely while seeing exactly the same information on their screens as those displayed on the Tempus IC2, providing much-needed peace of mind to the user and patient.
Connected medical support
Real-time access to support
Tempus IC2 allows intermittent users and non-medical professionals with basic medical training to obtain and transmit a patient's clinical vital signs data, including 12-lead ECG plus images and video. The device also offers a two-way voice connection to connect the user to professional medical support who can access a patient remotely.
Remote connection


Specifications
  • Physical dimensions
    Standalone size
    263mm x 216mm x 98mm (10.4" x 8.5" x 3.9")
    Standalone weight
    3 kg (6.6 lbs) including battery, RapidPakᵀᴹ clip, NBP cuff, hose, communications cable and headset
    Bag size
    401mm x 276mm x 154mm (15.8" x 10.9" x 6.1")
    Weight when packed in bag
    5.9 kg (13 lb) Tempus IC2 plus accessories (excluding power supply, and mains lead/battery charger)
  • Thermometer
    Type
    Handheld, Bluetooth-enabled infrared tympanic temperature
    Displayed range
    32–43°C (89.6–109.4°F) (target)
    Display resolution
    0.1°C (0.1°F)
    Accuracy
    ±0.2ºC (36–39°C) ±0.4°C (<lt/>36 <gt/>39°C) [±0.4ºF (96.8-102.2°F) ±0.5ºF (<lt/>96.8, <gt/>102.2°F)]
    Bluetooth communications range
    0–10 m (0–32'9")
  • 12-lead ECG
    Heart rate range
    30–300 bpm
    Input impedance
    <gt/>100 MΩ
    Dynamic range
    ±5 mV ac
    Accuracy
    ±3 %
    DC offset
    ±300 mV dc
    Frequency response
    0.05Hz–175Hz ±3 dB⁵
    Sample rate
    500 Hz
    Common mode rejection
    95 dB min, additional filters include mains, muscle, low and high pass and adaptive baseline zeroing
    Protection
    Protected against defibrillator charge at 5 kV
  • Capnometer - respiration rate
    Range
    0–149 bpm
    Accuracy
    0–70 bpm: ±1 bpm / 71–121 bpm: ±2 bpm / 122–149 bpm: ±3 bpm
  • Glucometer
    Test range (preset to mg/dL)
    10–600 mg/dL (0.6–33.3 mmol/L)
    Reading time
    3 seconds
    Bluetooth communications range
    Class 2, range 0–3 m (0–9' 10") in an open field
  • Pulse oximetry - pulse rate
    Range
    25-239 bpm
    Accuracy (all ages)
    No motion ≤3 digits; motion ≤5 digits
  • Pulse oximetry - SpO₂
    Range
    1–100 %
    Accuracy (adult/child)
    No motion/low perfusion: ±2 digits 70–100%; motion: ±3 digits, 70–100%
    Perfusion index range
    0.02–20%
  • Noninvasive blood pressure (NBP)
    Adult/large cuff range
    20–260 mmHg
    Adult/large cuff accuracy
    ±3 mmHg or ±2% (whichever is greater)
    Child cuff range
    20–260 mmHg
    Child cuff accuracy
    ±3 mmHg or ±2% (whichever is greater)
  • Battery and power
    Charger mains input voltage
    100–240 V
    Charger frequency
    50–60 Hz & 400 Hz
    Charger input currency
    0.9 A max at approximately 100 V
    Charger output voltage
    8.4 V dc
    Charger output current
    <lt/> 2.73 A
    Charger weight
    0.25 kg nominal
    Charger dimensions
    107mm x 67mm x 36.5mm (4.2" x 2.6" x 1.4")
    Power supply, external⁴
    PSU-rated 100–250 V, 50 Hz, 115 V 400 Hz
    Battery type
    Single, external, user-replaceable, lithium-ion
    Battery strength indicator
    Integral
    Battery life
    Min 11 hours¹ [default display brightness, SpO₂/EtCO₂ (25% of the time) and NBP every 15 minutes]
    Battery capacity
    Nominal 7.4 V, 10.2 Ah
    Battery shelf life
    Approximately 7 hours remaining after 1 year storage²
    Battery charge time from empty
    6³ hours
  • Integral digital camera
    Annotations
    Images can be annotated with text, colors, shapes which can be sent to Tempus IC2 by the hospital


Disclaimer
¹ Battery shelf life and run times are based on a new, fully charged battery stored and used at 20°C. Run time is based on RDT’s model of typical device usage in an incident.
² Subject to conditions of storage and use.
³ Only the RDT Battery Charger (part number: 01-1012) can be used with the Tempus IC2.
⁴ Battery may be charged (optionally) by the Tempus IC2 when running on mains power. Internal charging can be restricted if required.
⁵ Note that during monitoring prior and post recording ECG frequency response filters will be 0.5–175 Hz.